Experiment Data Role Blueprints (1)
Experiment Flow | consumed |
Modality | required |
maxOccurs | 1 |
Category: Method Description (optional, max. 1 occurrences)
Basic contextual information about the measurement method.
Parameter:
Method Name : String (optional, max. 1 occurrences)
Name of the method used for the measurement.
Parameter:
Method Reference : String (optional, max. 1 occurrences)
External reference to measurement method.
Category: Detector Properties (optional, max. 1 occurrences)
Parameter:
Flame Tip Type : String (optional, max. 1 occurrences)
Whether the flame tip is grounded or floating (=isolated).
Allowed values:
grounded, floating, other
Parameter:
Flame Tip Diameter (Length: [mm][in]) : Float (optional, max. 1 occurrences)
Diameter of the flame tip jet.
Parameter:
Detector Geometry : String (optional, max. 1 occurrences)
Collector electrode design.
Allowed values:
loop, tube, lateral plates, other
Parameter:
Detector Position : String (optional, max. 1 occurrences)
Position of the detector in the instrument. Often called 'A' or 'B', 'left' or 'right', ...
Parameter:
Make Up Gas Identity : String (optional, max. 1 occurrences)
Identity of the make up (also called auxiliary) gas of the detector.
Allowed values:
helium, hydrogen, nitrogen, argon, other
Category: Detector Settings (optional, max. 1 occurrences)
Parameter:
Block Temperature (Temperature: [K][°C][°F]) : Float (optional, max. 1 occurrences)
Configured temperature of the detector base block.
Parameter:
Bias Voltage (Voltage: [V]) : Float (optional, max. 1 occurrences)
DC polarizing voltage, typically between 100-300 V.
Parameter:
Hydrogen Flow Rate (Flow Rate: [mL/min][µL/min]) : Float (optional, max. 1 occurrences)
Flow rate of the hydrogen going to the flame.
Parameter:
Air Flow Rate (Flow Rate: [mL/min][µL/min]) : Float (optional, max. 1 occurrences)
Flow rate of the air going to the flame.
Parameter:
Oxygen Flow Rate (Flow Rate: [mL/min][µL/min]) : Float (optional, max. 1 occurrences)
Flow rate of the oxygen going to the flame.
Parameter:
Make Up Gas Flow Rate (Flow Rate: [mL/min][µL/min]) : Float (optional, max. 1 occurrences)
Flow rate of the make up (also called auxiliary) gas.
Category: Electrometer Settings (optional, max. 1 occurrences)
Parameter:
Range (Current: [A][mA][µA][nA][pA][fA]) : Float (optional, max. 1 occurrences)
Full-scale range of the electrometer, mostly given in full decades.
Parameter:
Current to Voltage Conversion Factor (Factor: [A/V]) : Float (optional, max. 1 occurrences)
Gain factor on the conversion from current to voltage.
Parameter:
Offset Current (Current: [µA][nA][pA][fA]) : Float (optional, max. 1 occurrences)
Bucking (zero adjustment) current to remove background ions, applied before current-to-voltage conversion.
Parameter:
Offset Voltage (Voltage: [V][mV][µV][nV]) : Float (optional, max. 1 occurrences)
Bucking (zero adjustment) voltage to remove background ions, applied after current-to-voltage conversion.
Parameter:
Attenuation Factor : Int (optional, max. 1 occurrences)
Pre-A/D conversion amplitude control factor, typically in the range of 1-1024.
Parameter:
Bandwidth (Rate: [µV/s][mV/s]) : Float (optional, max. 1 occurrences)
Response rate of the electrometer.
Parameter:
Time Constant (Time: [s][ms][µs]) : Float (optional, max. 1 occurrences)
Amount of time required for response to reach 1-1/e (~ 63.2 %) full scale. [IEEE 421A-1978]
Category: A/D Converter Settings (optional, max. 1 occurrences)
Parameter:
Channel ID : String (optional, max. 1 occurrences)
A/D converter channel ID, e.g. 'A', 'B', ... or '1', '2', ...
Parameter:
Mode : String (optional, max. 1 occurrences)
Acquisition scheme for detection.
Allowed values:
single-ended, differential
Parameter:
Range (Voltage: [V][mV][µV]) : Float (optional, max. 1 occurrences)
Maximum full-scale voltage of the A/D converter.
Parameter:
Range Polarity : String (optional, max. 1 occurrences)
Polarity of the voltage range of the A/D converter.
Allowed values:
positive, negative, bipolar
Parameter:
Resolution (Bits: [bits]) : Int (optional, max. 1 occurrences)
A/D converter signal resolution, given in number of bits, e.g. 12, 16, 24.
Parameter:
Digitization Rate (Frequency: [Hz]) : Float (optional, max. 1 occurrences)
Number of data points collected per second (after any averaging).
Parameter:
Averaging Factor (Count: ) : Float (optional, max. 1 occurrences)
Number of raw values averaged to obtain a data point.
Name | Time | Signal |
Documentation | TODO | Detector output signal |
Dependency | independent | dependent |
Data Type | Float | Float |
Modality | required | required |
maxOccurs | 1 | 1 |
Plot Scale | linear | linear |
Quantity:Units | Time
:
[ms][s][min] | Voltage
:
[V][mV][µV][nV][pV] Current : [A][mA][µA][nA][pA][fA] Arbitrary : [arbitrary] |
Allowed Values |